Appendix I
Calibraon cerficates of equipment used
}*"'#"
W
\*
H*@&ElJ
ilffiF":€-ffiAA
ETS-TESTCONSULT
LI M ITED
Performance Check /
C
alibration of Multinarameter
W
ter Oualitv Meter
Equipment
Ref.
No.
Model
No.
Date of Calibration
ET/EWO08/010
Manufacturer
Serial
No.
Caliblation Due
Date
YSI
Pro DSS
18E105421
3/312023
2/612023
Results
l Tempelature
(Method
Reference: Section
6 of
intelnation
Accreditation
New Zealand
Technical
Guide
no. 3 Second
edition March 2008
Thermometer
Calibration
Procedure)
Tolelance
Limit
("C):
+
2.0
2.pH
Reference:
APHA
19ed
4500-H*
B
Tolerance
Limit
(pH
unit):
+
0.10
3. Conductivity
Reference:
APHA
19ed 2510
Tolerance
Limit
(pS/cm):
+
10.0%
4. Salinity
Reference: APHA
19ed 2520
Tolerance Limit
(g/L):
+
10.0%
Tolerance
("C)
Displayed
Reading
("C)
Reading
of
Reference
Thel'mometer
('C)
-0.1
15.5
15.6
-0.1
25.0
25.1
-0.2
29.1
29.3
Tolerance
(pH
unit)
Displayed
Reading
(pH
unit)
Expected
Reading
(pH
unit)
4.00
6.86
9.18
Tolerance
(%)
Displayed
Reading
(prS/cm)
Expected
Reading
(pS/cm)
+0.4
145,8
145.2
+0.5
1421
1414
+0.7
12992
12892
-0.7
56343
s6761
Tolerance
(%)
Displayed
Reading
(g/L)
Expected
Reading
(g/L)
-0.6
9.94
10.0
-1.4
19.73
20.0
-0.1
29.97
30.0
Page 1 of2
\--*d4
I-.:ril}
*TKjffi#'l
\'*
R*ffi&'XIJ
ilRffi€EAA
ETS.TESTCONSULT
LI MITED
Performance Check
/ Calitrration of Multinarameter
W
ater Oualitv Meter
Equipment
Ref. No.
Model
No.
Date
of Calibration
ET/8W008/010
Manufacturer
Serial
No.
Calibration
Due Date
YSI
Pro DSS
1 8E1 05421
3/312023
216t2023
5.
Dissolved
Oxygen
(Method
Reference:
APHA
19ed
4500-0 G
Tolerance
Limit
(mg/L):
t 0.20
6.
Turbidity
Reference:
APHA
l9ed
2130 B)
Tolerance
Limit
(NTU):
t
10.0%
Tolerance
(mdl)
Displayed
Reading
(ndt')
Expected
Reading
(ne/L)
+0.11
1.84
1.73
-0.14
4.48
4.62
-0.08
5.83
5.91
Tolerance
(%)
Displayed
Reading
(NTU)
Expected
Reading
(NTU)
+3.0
10.3
l0
-0.5
39.8
40
+1.3
101.3
100
-0.97
396.1
400
The equipment
complies
n
I a""**"+*"roply
#
with the specified
requirements
and is deemed
acceptable
#
/
ur+aeeeptab+,+#
for use.
n
D.l.t. as
appropriate
Calibrated
by
ffir"
Approved
by
Page2
of2
\
,..'4
a.,.44'
!'...,4
\
H #
ffi
g,
P.l fiif,&€ f.1&'
F& A
frl
ETS.TESTCONSULT
LTD;
Performance
Gheck
of Water
Depth
Gauge
Equipment Ref.
No
Model No.
Date
of Check
Manufacturer
ETlEW10021012
SM-5
7691
6
Serial
No.
14-Jun-2023
15-Dec-2022
Due
Date
Speedtech
1
A rope was
marked of
lengths by
using a
calibrated
measuring tap
for every
1Ocm intervals.
2
lrhe initial
point
of the
rope
was connected
with a mass of
lead so that
it was lowered
into the
water
3
When the
rope reach
the bottom
of the
seabed,
read the mark
in the
rope to the
nearest Scm.
4
Use the
water depth
gauge
to
measure
the result
at the same
position.
5
Repeat measuring
for total 3 times
and
use the
mean value to
calculate
the
percentage
error.
6
3 different
measuring
should be
check
to cover the
depth
for normal operation.
Method
Calibration
Due Date
24-Jun-2023
Calibrated
Measuring
Tape
Equipment
Ref. No. ETt2010105
Range
(m)
Reading
by
Measuring
Tape
(m),
Dr
Reading
by Water
Depth
Gauge
(m),
Dx
Percentage
Error
(%)
^
Result
#
0-5m
3.75 Mean
3.7
Mean
2.1%
Accept
3.90
3.88
3.8
3.8
4.00
3.9
5-10m
7.30
Mean
7.3
Mean
0.6%
Accept
7.85
7.78
7.9
7.83
8.20
8.3
10
-
20m
12.55
Mean
12.7
Mean
1.3%
Accept
12.75
12.90
12.9
13.07
13.40
13.6
^
Percentage
Error,
(%)
=
(
|
Dx
-
Dr
| )
/
Dr
)
x
100%
Resu/fs
Acceptance
Criteria
# Result
accept
if Percentage
Error
<
10o/o, otherwise
reject.
The
Water Depth
Gauge complies
*
/4€e€-n€+-€€'mp{y
*
with the
specified
requirements
and
is deemed
acceptable
*
/-unaeeeptable-*
for
use.
*
Delete as appropriate
Approved
by
Checked by
6--!-/
*\:.,';S-
[-,*f#
\*#
\*".
***4rxll
ilffirda'RAa
ETS-TESTCONSULT
LI M ITED
Form
E/CE/L/l 5/lssue 2 vl
lnternal
Calibration &
Performance
Check of
pH
Meter
Equipment
Ref. No
Model No.
Date of Calibration
Manufacturer
Serial No.
Calibration
Due Date
HANNA
lnstruments
ETtEW1007l011
J0046801
Ht 9125
61312023
5t4t2023
Liquid
Junction
Error
Primary Standard
Solution
Used
Temperature
of Solution
:
pH
value of diluted
buffer :
(Observed
Deviation)
0.005 / 0.010
Liquid
Junction
Error
(ApH;)
=
ApH
-
LpH,
=
Ref
No. of Primary
Solution:
003/5.3/8
PH
(S)
=
9.225
/ 9.180
ApH
=
pH(S)- pH
of
diluted buffer
=
0.005
/ 0.020
Borax
9.22
t9.20
19.9
125.1
LPH
u=
0.01
Shift
on Stirring
pH
of buffer
solution
(with
stirring),
pH"
=
Shift
on stirring,
ApH"=
pH"
-
pH(S)
-
ApH,
=
9.21 t9.18
0.015 /
0.000
Noise
Noise, ApHn
=
difference
between max
and min reading
0.01 / 0.01
Ref. No. of reference
thermometer
used:
Temperature
record
from the reference
thermometer
(Tp)
Temperature
record from
the ATC
(T416):
Temperature
Difference,
lt*
-
tor"
I
Correction
C
c
c
C
Verification
of ATC
25.2
20.2 I
25.1
20.1 I
0.1 I
0.'1
0.1 I
0.1
o
ETtO521t025 I
ET105211030
o
o
o
Performance Characteristic
Acceptable
Range
Liquid Junction
Error
^pHj
<0.05
Shift on
Stirring
ApHs
<0.02
Noise
ApHn
<0.02
Verifcation of ATC
Temperatu
re Difference
<0.50c
Acceptance
Criteria
The
pH
meter complies
*
/ d€es
not €€mply
*
with the specified
requirements
and is deemed
acceptable
*
/
{+n€€€eptable
"
for use. Measurements
are traceable
to national
standards.
*
Delete
as appropriate
Checked
by
Calibrated
by
?{
cPE/01
!=*=;r'l
].'*ter:'i
\.-**
H*FSXlI
ilFH€.FAA
ETS.TESTCONSULT
LI
M ITED
Fom
E/CE/L/15/lssue 2
lnternal
Calibration
& Performance
Check
of
pH
Meter
Equipment
Ref.
No.
Model
No.
Date of Calibration
Manufacturer
Serial
No.
Calibration Due
Date
HANNA
lnstruments
ETtEW|OOTl0ll
J0046801
Ht 9125
6t3t2023
7t212023
Liquid
Junction
Error
Primary
Standard
Solution
Used
Temperature
of Solution
:
pH
value of diluted
buffer
:
(Observed
Deviation)
Liquid Junction
Error
(lpHi)
=
ApH
-
LgHu=
0.005
/ 0.000
Ref
No. of
Primary Solution:
00315.317
PH
(S)
=
9.225 / 9.180
ApH=pH(S)-pH
of
diluted buffer
=
0.005
/ 0.010
Borax
19.8
t24.9
9.2319.19
LgH
u=
0.01
Shift
on Stirring
pH
of buffer
solution
(with
stirring),
PH,
=
Shift
on stirring,
AFH,=
pH"
-
pH(S)
- ApH,
=
9.22 t9.18
0.010
/ 0.000
Noise
Noise,
ApHn
=
difference between
max and
min reading
0.00
/ 0.02
Ref.
No.
of
reference
thermometer
used:
Temperature
record
from the
reference thermometer
(Tp)
Temperature
record
from the
ATC
(Ta16):
Temperature
Difference,
lT*
-
Tor"
I
Correction
c
c
c
c
Verification
of ATC
20.2
I
25.1
24.9
20.3
I
0.1
I
0.2
0.1 I
o
0.2
ETt0521t025
I ETl0s211030
o
o
Performance
Characteristic
Acceptable
Ranqe
Liquid
Junction
Error
ApHj
<0.05
Shift
on Stirrins
ApHs
<0.02
Noise
ApHn
<0.02
Verifcation
of ATC
Temperature
Difference
<0.50c
Acceptance
Criteria
The
pH
meter complies
*
/ d€es n€t
€emply
*
with the specified
requirements
and is deemed
acceptable
"
/
un€€€eptable
*
for use.
Measurements
are
traceable to
national standards.
*
Delete as appropriate
Checked
by
Calibrated
by
W