,\**M
reffi H*@gEll il&16€'FAfr
ETS-TESTCONSULT LIMITED
Performance Check / Calibration of Multinarameter Water Oualitv Meter
Equiprnent Ref. No.
Model No.
Date of Calibration
ET/EW008/010 Manufactuler
Serial No.
Calibration Due Date
YSI
Pro DSS 18E105421
21612023 l/912023
Results
1. Temperatule
(Method Reference: Section 6 of intelnation Accleditation New Zealand Technical Guide no. 3 Second edition March 2008
Working Thermometel Cal ibration
Tolerance Limit ("C): + 2.0
2,pH
Reference: APFIA 19ed 4500-H+ B)
Tolerance Limit (pH unit): + 0.10
3. Conductivity
Reference: APFIA 19ed 2510 B)
Tolerance Limit (pS/cm): + 10.0%
4. Salinity
(Method Reference: APHA l9ed 2520 B)
Expected Reading (g/L) Displayed Reading (g/L) Toletance (o/o)
10.0 9.97 -0.3
20.0 19.68 -1.6
30.0 30.04 0.1
Tolelance Limit (g/L): + 10.0%
Tolerance (oC)
Reading of Reference Thermometer' (oC) Displayed Reading ("C)
0.1
15.6 15.7
25.2 0.1
2s.1
-0.1
29.3 29.2
Tolerance (pl-l unit)
Expected Readine (pH unit) Displayed Reading (pH unit)
4.00
6.86
9.1 8
Displayed Reading (pS/cm) Tolemnce (7o)
Expected Reading (pS/cm)
145.6 +0.3
145.2
+0.6
1414 1423
+0.8
12892 13004
56393 -0.6
56761
Page I of2