APPENDIX C
Calibration certificates of equipment
,\**M
reffi H*@gEll il&16€'FAfr
ETS-TESTCONSULT LIMITED
Performance Check / Calibration of Multinarameter Water Oualitv Meter
Equiprnent Ref. No.
Model No.
Date of Calibration
ET/EW008/010 Manufactuler
Serial No.
Calibration Due Date
YSI
Pro DSS 18E105421
21612023 l/912023
Results
1. Temperatule
(Method Reference: Section 6 of intelnation Accleditation New Zealand Technical Guide no. 3 Second edition March 2008
Working Thermometel Cal ibration
Tolerance Limit ("C): + 2.0
2,pH
Reference: APFIA 19ed 4500-H+ B)
Tolerance Limit (pH unit): + 0.10
3. Conductivity
Reference: APFIA 19ed 2510 B)
Tolerance Limit (pS/cm): + 10.0%
4. Salinity
(Method Reference: APHA l9ed 2520 B)
Expected Reading (g/L) Displayed Reading (g/L) Toletance (o/o)
10.0 9.97 -0.3
20.0 19.68 -1.6
30.0 30.04 0.1
Tolelance Limit (g/L): + 10.0%
Tolerance (oC)
Reading of Reference Thermometer' (oC) Displayed Reading ("C)
0.1
15.6 15.7
25.2 0.1
2s.1
-0.1
29.3 29.2
Tolerance (pl-l unit)
Expected Readine (pH unit) Displayed Reading (pH unit)
4.00
6.86
9.1 8
Displayed Reading (pS/cm) Tolemnce (7o)
Expected Reading (pS/cm)
145.6 +0.3
145.2
+0.6
1414 1423
+0.8
12892 13004
56393 -0.6
56761
Page I of2
#sHtl ilffiffi€Eaaa
TESTCONSULT LIMITED
R$6
ETS
L.*W
t=;#a
*W
Performance Check / Calibration of Multinarameter Water Oualitv Meter
Equipment Ref. No.
Model No.
Date of Calibration
Manufacturer
Serial No.
Calibration Due Date
ET/EW/008/010
Pro DSS
2/6/2023 U9t2023
YSI
I 8E1 05421
5. Dissolved Oxygen
Reference: APHA 19ed 4500-0 G)
Tolerance Limit (mgll-): + 0.20
6. Turbidity
(Method Reference: APHA l9ed 2130 B
Tolerance Limit (NTU): x10.0%
Displayed Reading (ms./L) Tolerance (me/L)
Expected Reading (mg/L)
+0.08
1.73 1.81
4.49 -0.1 3
4.62
-0.0s
s.91 5.86
Tolerance (o/o)
Expected Readins (NTU) Displayed Reading (NTU)
10.2 +2.0
10
+1.75
40 40.7
102.5 +2.5
100
-0.70
400 397.2
The equipment complies # / dees-ne+eomply # with the specified requirements and is deemed acceptable # / una€€€ptabl.e-# for use.
u Delet. as appropriate
Calibrated by Approved by
/.r 6
0
l\4anaq p u'
(Ll
*J Page2 of2