
,\**M
reffi
H*@gEll il&16€'FAfr
ETS-TESTCONSULT
LIMITED
Performance Check / Calibration
of M
ultinarameter Water Oualitv Meter
Equiprnent
Ref. No.
Model
No.
Date
of Calibration
ET/EW008/010
Manufactuler
Serial No.
Calibration
Due Date
YSI
Pro DSS
18E105421
21612023
l/912023
Results
1. Temperatule
(Method
Reference: Section
6 of intelnation
Accleditation
New Zealand Technical Guide
no. 3 Second
edition March
2008
Working
Thermometel
Cal ibration
Tolerance
Limit
("C):
+
2.0
2,pH
Reference: APFIA
19ed 4500-H+ B)
Tolerance
Limit
(pH
unit):
+
0.10
3. Conductivity
Reference:
APFIA 19ed 2510 B)
Tolerance
Limit
(pS/cm):
+
10.0%
4. Salinity
(Method
Reference: APHA
l9ed 2520 B)
Expected
Reading
(g/L)
Displayed
Reading
(g/L)
Toletance
(o/o)
10.0
9.97
-0.3
20.0
19.68
-1.6
30.0
30.04
0.1
Tolelance
Limit
(g/L):
+
10.0%
Tolerance
(oC)
Reading
of Reference
Thermometer'
(oC)
Displayed Reading
("C)
0.1
15.6
15.7
25.2
0.1
2s.1
-0.1
29.3
29.2
Tolerance
(pl-l
unit)
Expected Readine
(pH
unit) Displayed
Reading
(pH
unit)
4.00
6.86
9.1 8
Displayed Reading
(pS/cm)
Tolemnce
(7o)
Expected Reading
(pS/cm)
145.6
+0.3
145.2
+0.6
1414
1423
+0.8
12892
13004
56393
-0.6
56761
Page
I
of2