-".#W
\*""d
*\**
H#e&ilsmH€RAA
ETS.TESTCONSULT
LI MITED
ration of Multinarameter Water
Performance Check / Calib
Oualitv Meter
Equipment
Ref. No
Model No.
Date
of Calibration
Manufacturer
Serial
No.
Calibration
Due
Date
ET/EW008/010
Pro DSS
2/612023
YSI
I 8E105421
1/912023
5.
Dissolved
Oxygen
(Method
Refereuce:
APHA
19ed 4500-O
G)
Expected
Reading
(np.lL)
Displayed
Reading
(mg/L)
Tolerance
(rne/L)
1.73
I .81
+0.08
4.62
4.49
-0.1
3
5.91
s.86
-0.05
Tolerance
Limit
(mgil):
+
0.20
6. Turbidity
Reference:
APHA 19ed
2130 B)
Tolerance
Limit
(NTU):
t 10.0%
Tolerance
(%o)
Displayed
Readins
(NTU)
Expected
Reading
(NTU)
+2.0
10
10.2
+1.75
40.7
40
+2.5
102.5
100
-0.70
397.2
400
The equipment
cornplies
n
I a""s"o+*"*pty
#
with
the specified
requirements
and
is deemed
acceptable
#
/
unaeeeptabl+#
for use
o
D.l.t.
as appropriate
Calibrated
by
Approved
by
Page